L2 - R4.6 Dispensing Warning

Modified on Mon, 26 May at 11:23 PM

According to the problem analysis result done by CCA, you can directly begin the subsequent troubleshooting process from the corresponding potential root cause.



 1. BiB

STEP 1 - Verify if the BiB is not empty or folded.

  • Result 1 - The BiB is not empty or folded. 
    Start "Pump" chapter.


  • Result 2 - The BiB is empty or folded
    Start STEP 2.

STEP 2 - Change BiB / Push the remaining ingredients in the BiB towards to the spout, and fill tube to see if the ingredients can be dispensed smoothly.

  • Result 1 - The ingredients can be dispensed smoothly. 
    Start STEP 3.
    Root cause: Empty BiB / Folded BiB.


  • Result 2 - Still dispense warning
    Start "Pump" chapter.


 2. Pump  


STEP 1 - Activate the malfunctioning pump, and measure the voltage at the control board connector for the corresponding pump. The expected reading should be 24V

  • Result 1 - The reading is 24V. Control board functioning properly, continue diagnosing pump failure. 
    Start STEP 3.


  • Result 2 - The reading is not 24V. Control board defective, replace control board and re-test.
    Start STEP 2.


STEP 2 - Replace the control board (see SOP here) and verify pump functionality.

  • Result 1 - The pump works fine. 
    Start STEP 6.
    Root cause: The control board failed.

  • Result 2 - The pump is still not working.
    Start STEP 3.

STEP 3 Replace the pump (see fridge pump SOP / ambient pump SOP here) and verify pump functionality.

  • Result 1 - The pump works fine. 
    Start STEP 6.
    Root cause: The pump failed.

  • Result 2 - The pump is still not working 
    Start STEP 4.

STEP 4 - Inspect the pipeline for leaks that may reduce pump extraction efficiency.

  • Result 1 - There are leaks. 
    Start STEP 5.

  • Result 2 - There are no leaks. 
    Start "Flowmeter" chapter.


STEP 5 - Seal the leaks as follows, make sure no leaks, and start STEP 6.

Pipeline rupture: Replace the damaged section of the pipeline.

Incomplete connection sealing: Properly install the pipeline and clamps to their designated positions (following the Standard Operating Procedure - SOP).


STEP 6 - Make a drink to verify the resolution of the dispensing warning issue.

  • Result 1 - No dispense warning pops up. 
    Issue close. (If the issue is closed after completing STEP 5, the root cause for your case is tubing leaks)

  • Result 2 - Dispensing warning still happens.
    Start "Flowmeter" chapter.





 3. Flowmeter  


STEP 1 - Flush the line with hot water and check if the flowmeter can read.

  • Result 1 - The flowmeter is readable now.
    Start 
    STEP 4.
    Root cause: the flowmeter clog.

  • Result 2 - The flowmeter is still unreadable.
    Start STEP 2

STEP 2 - Replace the flowmeter (see fridge flowmeter SOP / ambient flowmeter SOP here), and verify flowmeter functionality.

  • Result 1 - The flowmeter is readable now.
    Start 
    STEP 4.
    Root cause: the flowmeter fail.

  • Result 2 - The flowmeter is still unreadable.
    Start STEP 3.

STEP 3 - Replace the control board (see SOP here), and verify flowmeter functionality.

  • Result 1 - The flowmeter is readable now.
    Start 
    STEP 4.
    Root cause: the control board failed

  • Result 2 - The flowmeter is still unreadable.  
    Escalate to PSE.

STEP 4 - Fill tube, and make a drink to verify the resolution of the dispensing warning issue.

  • Result 1 - No dispense warning pops up. 
    Issue close.

  • Result 2 - Dispensing warning still happens.
    Escalate to PSE. 









Return to index

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article