According to the problem analysis result done by CCA, you can directly begin the subsequent troubleshooting process from the corresponding potential root cause.
1. Blender
STEP 1 - Put the blender cup on the blender and check if the blender can blend typically.
- Result 1 - Can blend.
Start "Power" chapter.
- Result 2 - Cannot blend.
Start STEP 2.
STEP 2 - Follow the steps in the blender failure troubleshooting guide to fix the blender. Then, test if the cleaning tool can be detected.
- Result 1 - Can be detected.
Issue close.
Root cause: The blender fails - Result 2 - Cannot be detected.
Start "Power" chapter.
2. Power
STEP 1 - Reconnect the wires with the electrodes, then verify if the voltage on the electrodes is 24V DC.
- Result 1 - The reading is 24V.
Start "Clenaing Tool" chapter. - Result 2 - The reading is not 24V.
Start STEP 2
STEP 2 - Verify that the cleaning tool wire on the control board is securely connected.
- Result 1 - Not securely connected.
Start STEP 3. - Result 2 - Securely connected.
Start STEP 4.
STEP 3 - Reconnect the wire to the control board. Then, verify the cleaning tool detection when starting the cleaning process.
- Result 1 - Can be detected.
Issue closed.
Root cause: Wire disconnected. - Result 2 - Cannot be detected.
Start STEP 4.
STEP 4 - Measure the voltage at the cleaning tool wire connector on the control board. The expected reading should be 24V DC.
- Result 1 - The reading is 24V.
Start "Clenaing Tool" chapter. - Result 2 - The reading is not 24V.
Start STEP 5.
1-3-3-11 - Replace the control board (see SOP here). Then, verify the cleaning tool detection when starting the cleaning process.
- Result 1 - Can be detected.
Issue close.
Root cause: The control board fails. - Result 2 - Cannot be detected.
Start "Clenaing Tool" chapter.
3. Cleaning Tool
STEP 1 - Verify if the spring-loaded pins under the cleaning tool can rebound properly and are well-connected with the cleaning tool ball valve wires.
- Result 1 - The spring-loaded pins can rebound, and the wires are well connected.
Escalate to PSE. - Result 2 - The spring-loaded pins cannot rebound.
Start STEP 2.
- Result 3 - The wires are not well connected to the pins.
Start STEP 3.
STEP 2 - Replace the spring, and verify the cleaning tool detection when starting the cleaning process.
- Result 1 - Can be detected.
Issue close.
Root cause: Spring on the pin malfunction. - Result 2 - Cannot be detected.
Restart STEP 1.
STEP 3 - Reconnect the wires, and verify the cleaning tool detection when starting the cleaning process.
- Result 1 - Can be detected.
Issue close.
Root cause: Cleaning tool ball valve wires disconnected. - Result 2 - Cannot be detected.
Restart STEP 1.
Was this article helpful?
That’s Great!
Thank you for your feedback
Sorry! We couldn't be helpful
Thank you for your feedback
Feedback sent
We appreciate your effort and will try to fix the article